PubMed · 2003 · 152 citations · 3 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic Nuclear MedicineDiagnostic RadiologyContrast AgentsContrast MediaGastrointestinal ImagingGastrointestinal RadiologyMedical ImagingPositron Emission TomographyContrastContrast-enhanced Computed TomographyHounsfield Unit MeasurementHounsfield UnitsOral Contrast MediaPet Attenuation CorrectionPet-computed TomographyTracer Specific Activity
UNLABELLED: The aims of the study were to evaluate the effects of oral contrast on apparent tracer activity measured with PET/CT when using CT attenuation correction and to report our initial experience in the use of oral contrast with PET/CT. METHODS: Phantom studies with (18)F activity and saline bags or syringes filled with barium or gastrografin of varying densities were performed using a PET/CT scanner (CT attenuation correction). In the study, 91 clinical patients received dilute oral contrast and were evaluated by whole-body (18)F-FDG PET. RESULTS: A phantom experiment with CT contrast (1.3% weight/volume [w/v] barium) showed a "cold" area in the cold stomach whereas a phantom with high-density barium (98% w/v) showed an artifactual focus of intense "activity" in the cold stomach. In clinical studies, stomach and right colon were opacified by CT contrast. Maximal measured contrast density was 239 Hounsfield units. CONCLUSION: High-density barium causes overestimation of tissue (18)F-FDG concentration. Low-density barium does not cause significant artifacts and appears suitable for clinical use.
3
Dilute barium as a contrast agent for abdominal CT
M A Bosniak · American Journal of Roentgenology · 1980 · 41 citations · Full text