Publication | Closed Access
Characterization of an image-plate detector used for quantitative small-angle-scattering studies
37
Citations
0
References
1993
Year
EngineeringMeasurementMicroscopyOptical TestingImage AnalysisImage PlateImage-plate Readout SystemComputational ImagingInstrumentationDetection TechnologyNuclear MedicineRadiologyHealth SciencesRadiation DetectionMedical ImagingPhysicsInexpensive Neutron-imaging SystemNeutron SourceRadiographic ImagingMicroscope Image ProcessingElectronic ImagingImage-plate DetectorNeutron Scattering
This paper describes the characterization of an image-plate readout system used to calculate small-angle-scattering intensities as cross sections per unit volume. The raw data are the latent phosphorescent excitations stored on an image plate. Self-decay, self exposure, resolution, smearing, dynamic range and sensitivity are measured. Preliminary experiments show that an inexpensive neutron-imaging system can be obtained when the plates are placed behind a suitable neutron-to-X-ray converter.