SAMPLING OF GASEOUS POLLUTANTS ON ACTIVATED CHARCOAL WITH 900 mg TUBES

J. P. Guenier, Jean‐Pierre Müller

The Annals of Occupational Hygiene · 1984 · 26 citations · 0 references

Concepts

Abstract

Sampling of organic vapours on activated charcoal is a widely used technique. INRS has been supplying 900 mg tubes, suitable for a very wide scale of concentrations and for many compounds, even the most volatile. This paper reports experiments testing the validity of this equipment, and a few points, which are seldom discussed, are also examined. The breakthrough volume, proportional to the amount of charcoal and independent of the flow rate (0.5–1.5 1. min −1 ), is different for each compound and is smaller when the concentration to be measured is lower. When two pollutants are collected simultaneously, preferential fixation occurs, in some cases reducing the charcoal adsorption capacity. Stability of the loaded tubes proves generally satisfactory. After desorption, the conservation of the solutions should be checked, particularly for low concentration oxygenated compounds. In the Appendix, some results of laboratory tests are reported concerning most compounds of various chemical functions.