EngineeringBioengineeringElectrical EngineeringBiomedical Signal ProcessingElectrophysiologyInstrumentationBiomedical InstrumentationBiopotential MeasurementElectrical MeasurementElectrocardiogram AnalysisElectrocardiographyElectrode Placement12-Lead ElectrocardiographyElectrode - Skin InterfaceElectrode Array PlacementElectrode Impedance CharacterizationElectrode Placement AccuracyElectrode-skin InterfaceSurface ElectrocardiographySurface Impedance
In electrocardiographic recording one of the main sources of error is an imperfect mechanical contact in the skin-electrode interface because it causes high impedances that can contribute to distortion in the ECG. We measured contact impedance to determine the corresponding values to a reliable contact in this interface in 10 electrodes of the standard 12-lead ECG of 5 normal subjects from 1 Hz to 50 kHz. An impedance decrease for increasing frequency and a wider dispersion in low frequencies was observed. In the ECG frequency range, maximum contact impedances varied from 20 k/spl Omega/ in 50 Hz and 100 Hz to 75 k/spl Omega/ in 1 Hz. Because we are interested in analyzing ECG low frequency waves, we use the range of impedance values in 5 Hz for each electrode to detect a poor electrode placement in the 12-lead ECG recording.
8
Medical Instrumentation-Application and Design
John G. Webster · Journal of Clinical Engineering · 1978 · 1.8K citations
A new system of multiple-lead exercise electrocardiography
Robert E. Mason, I.N. Likar · American Heart Journal · 1966 · 508 citations
Medicine, Cardiology, Physiology +14
Skin impedance from 1 Hz to 1 MHz
J. Rosell, J. Colominas, Pere J. Riu et al. · IEEE Transactions on Biomedical Engineering · 1988 · 341 citations
Engineering, Biomedical Engineering, Electrical Engineering +10
Skin-Electrode Impedance and Its Effect on Recording Cardiac Potentials
Madison S. Spach, Roger C. Barr, James W. Havstad et al. · Circulation · 1966 · 109 citations