Publication | Open Access
Report of the IFCC Working Group for Standardization of Thyroid Function Tests; Part 3: Total Thyroxine and Total Triiodothyronine
94
Citations
35
References
2010
Year
The study showed that most assays had bias largely correctable by establishing calibration traceability to a cRMP and that the majority performed well. Some assays, however, would benefit from improved precision, within-run stability, and between-run consistency.