A high-resolution fast spin-echo inversion-recovery sequence for preoperative localization of the internal globus pallidus.

Caroline A. Reich, Patricia A. Hudgins, Scott K. Sheppard, Philip A. Starr, Roy A.E. Bakay

PubMed · 2000 · 29 citations · 11 references

Full text

Open access

Concepts

Abstract

A fast spin-echo inversion-recovery (FSE-IR) sequence is described for its utility regarding surgical planning for patients with Parkinson's disease (PD) who are undergoing microelectrode-guided internal globus pallidus (GPi) ablation. Images from thirty-seven adult patients with PD were reviewed and visualization of the GPi, globus pallidus externa (GPe), and the intervening lamina was noted. High-resolution images were acquired from all patients despite the external hardware and the patients' movement disorder. In all cases, the conventional surgical trajectory, determined indirectly by a fixed measurement from the anteroposterior commissure line, was modified by the ability to visualize the GPi and optic tract directly. This sequence facilitated accurate stereotactic targeting.

References