Whole cortex, 64 channel SQUID biomagnetometer system

Jan Vrba, K. Betts, M. B. Burbank, Teresa Cheung, A. A. Fife, G. Haid, P. R. Kubik, S. Lee, J. Adam McCubbin, Joseph C. McKay,

IEEE Transactions on Applied Superconductivity · 1993 · 67 citations · 6 references

Concepts

Abstract

The authors report on the development and testing of a novel, whole cortex, 64-channel superconducting quantum interference device (SQUID) biomagnetometer system operating in an unshielded environment. The essential features of this instrument, including the cryogenics system, the room-temperature digital electronics, and signal processing capabilities, are described. A noise cancellation scheme is incorporated which allows extraction of biomagnetic signals over the surface of the cortex while operating in the presence of large background magnetic interference. A system description and the results of preliminary magneto-encephalographic measurements are presented. Measurements of noise performance and human auditory evoked response in an unshielded industrial environment have largely validated the noise cancellation methods incorporated in the instrument.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

6