PharmacyChemistryAnalytical ChemistryElectrochemistrySensorsAnalytical InstrumentationElectroanalytical ChemistryElectrochemical DetectionElectrochemical SensorsIon-selective ElectrodesReference ElectrodesDisposable ElectrodesElectrode LifetimeExternal Reference ElectrodesIon Selective SensorsIon-selective Electrode SensorsIon-selective Reference ElectrodesPotential DifferenceReference Electrode Design
A disposable reference electrode was developed which can be used for environmental and medical analysis. It is an Ag/AgCl electrode with an integrated junction, prepared using inexpensive materials such as glass-fibre filter medium and heat-sealing film with screen printing and encapsulating by lamination. The potential difference between this electrode and a commercial electrode is hardly influenced by concentration changes and pH (3–11) of the analyte solution. The shelf-life and the operational lifetime of the reference electrode were investigated. Applications with disposable ion selective sensors are demonstrated.
5
Disposable ion-selective electrodes
M. Borchardt, C. Dumschat, Karl Cammann et al. · Sensors and Actuators B Chemical · 1995 · 28 citations
Disposable reference electrode
Christoph Diekmann, C. Dumschat, Karl Cammann et al. · Sensors and Actuators B Chemical · 1995 · 20 citations
C. Dumschat, M. Borchardt, Christoph Diekmann et al. · Sensors and Actuators B Chemical · 1995 · 20 citations
M. Borchardt, Christoph Diekmann, C. Dumschat et al. · Talanta · 1994 · 17 citations