Acta Radiologica · 1997 · 14 citations · 8 references
Veterinary MedicineOrthopedic SurgeryRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography PhysicsImage Quality AssessmentMedical ImagingRadiographic ImagingScanning TechniquesClinical Image QualityComputed Tomography Dose MeasurementConventional RadiographyLow-dose ReconstructionMusculoskeletal Computed TomographyRadiation Dose OptimizationRadiographic Image QualityReconstruction Algorithms
PURPOSE: The aim was to select an optimal technique for low-dose high-resolution CT of the sacroiliac joints (SJ). MATERIAL AND METHODS: Dose measurements were performed on a Rando anthropomorphic phantom using thermoluminescence dosimeters for 4 CT protocols and 2 conventional radiography protocols used for SJ evaluation. Six available reconstruction algorithms were tested on CT protocols using 285-665 mAs and 120 or 130 kVp settings and noncontiguous 1.5-mm-thin sections with 3.5-mm intervals. Settings with optimum performance on phantom tests were also applied in a series of 10 patients with SJ arthropathies. RESULTS: A CT protocol using 120 kVp/175 mA/2.9 s/1.5-mm slice thickness/5-mm table increment implied the lower radiation dose among all examination protocols tested and provided high image quality of the SJ. A reconstruction algorithm yielding images of improved spatial resolution with acceptable noise was selected. CONCLUSION: A high spatial frequency reconstruction algorithm, and 120 kVp and 508 mAs were considered optimal for a low-dose CT examination of the SJ that employed narrow (1.5 mm) slice images with interspacing.
8
The normal sacroiliac joint: a CT study of asymptomatic patients.
J B Vogler, W. H. Brown, C A Helms et al. · Radiology · 1984 · 145 citations
Thomas L. Lawson, W. Dennis Foley, Guillermo F. Carrera et al. · Journal of Computer Assisted Tomography · 1982 · 89 citations
Radiation exposure of patients who undergo CT of the trunk.
R. Mini, Peter Vock, R Mury et al. · Radiology · 1995 · 61 citations