IEEE Transactions on Instrumentation and Measurement · 1999 · 28 citations · 2 references
EngineeringElectrical EngineeringMetrologyMicroelectronicsUncertainty QuantificationCurrent Measurement InstrumentsElectrical MeasurementEquipment CalibrationMeasurement ErrorMeasurement UncertaintyOhm's LawCurrent Measurement AccuracyCurrent Measurement UncertaintyHigh-current MeasurementHigh-precision MeasurementPrecision MeasurementResistance Uncertainty AnalysisResistorsShunt Measurement
The characteristics of laboratory resistors and high-precision resistors in the principal range, 1 /spl Omega/ to 10 k/spl Omega/, are reviewed. Also reviewed is the history of the shunt and of shunt construction. A method is given for improving the uncertainty of shunts by a factor of five by measuring temperature and providing the resistance-temperature characteristic. Data for high-current shunts using this method are included showing the improvement even in cases where shunt behavior is less than expected. Data are also given for the uncertainty improvement achieved with a common meter type, high-current shunt. The results of efforts to identify the cause of the behavior that went wrong in high-precision shunts are also presented.
2
Forest K. Harris, W. G. Wadey · American Journal of Physics · 1952 · 122 citations