Journal of Applied Clinical Medical Physics · 2011 · 59 citations · 7 references
MedicineNeurosurgeryRadiation OncologyBiomedical ImagingNeuroradiologyTreatment PlanningDose PrescriptionStereotactic NeurosurgeryStereotactic RadiosurgeryStereotactic RadiotherapyStereotactic SurgeryTarget Volume DelineationConformity AssessmentDose Conformity AssessmentDose Volume HistogramDose-volume Histogram AnalysisStereotactic Dose DistributionStereotactic Radiosurgery DosimetryTarget Volume Evaluation MetricsTreatment Plan Evaluation
A database of clinically approved stereotactic radiosurgery treatment plans was created. One hundred and seventy targets in the database were then retrospectively evaluated using conformity indices suggested by RTOG, SALT-Lomax and Paddick. Relationships between the three alternative conformity indices were determined. The Paddick index combines the information provided by the RTOG and SALT-Lomax indices into a single index. The variation in the geometric overlap ratio, which is related to the SALT-Lomax index, was found to be not clinically relevant for our cohort of patients, and thus the Paddick and RTOG indices can be directly related. It was found that access to a dose volume histogram or dose distribution for a treatment plan renders the RTOG conformity index sufficient for plan quality evaluation.
7