IEEE Transactions on Biomedical Engineering · 1973 · 39 citations · 15 references
MedicineCardiologyVeterinary PhysiologyCardiothoracic SurgeryHemodynamic ManagementInterventional CardiologyBlood Pressure ControlBlood Pressure ManagementBlood Pressure RegulationIntra-aortic Balloon PumpingOperating Point DeterminationOperating Point OptimizationSteepest Descent
A closed-loop control scheme for the control of intra-aortic balloon pumping has been developed and tested in dog experiments. A performance index reflecting the general objectives of balloon-assist pumping is developed and a modified steepest ascent control algorithm is utilized for the selection of a proper operating point for the balloon during its pumping cycle. This paper attempts to indicate the feasibility of closed-loop control of balloon pumping, and particularly its flexibility in achieving both diastolic augmentation of mean aortic pressure and control of the level of end-diastolic pressure (EDP)-an important factor in reducing heart work.
15
R. E. Kalman, Douglass J. Wilde · American Mathematical Monthly · 1965 · 216 citations
Velocity patterns in the aorta
W. A. Seed, Nigel B. Wood · Cardiovascular Research · 1971 · 200 citations
Use of Fourier Series for the Analysis of Biological Systems
Ernst O. Attinger, Alexis Anne, D. A. McDonald · Biophysical Journal · 1966 · 144 citations · Full text
Engineering, Biology, Biophysics +14