On the Feasibility of Closed-Loop Control Of Intra-Aortic Balloon Pumping

John W. Clark, Gerald R. Kane, H. M. Bourland

IEEE Transactions on Biomedical Engineering · 1973 · 39 citations · 15 references

Concepts

Abstract

A closed-loop control scheme for the control of intra-aortic balloon pumping has been developed and tested in dog experiments. A performance index reflecting the general objectives of balloon-assist pumping is developed and a modified steepest ascent control algorithm is utilized for the selection of a proper operating point for the balloon during its pumping cycle. This paper attempts to indicate the feasibility of closed-loop control of balloon pumping, and particularly its flexibility in achieving both diastolic augmentation of mean aortic pressure and control of the level of end-diastolic pressure (EDP)-an important factor in reducing heart work.

References

15