2007 · 23 citations · 7 references
EngineeringBiomedical EngineeringElectrical EngineeringBioinstrumentationMicroelectronicsSystem-on-chip DesignCmos Circuit DesignDevice DesignImplantable Biomedical DevicesImplantable BiosensorsImplantable MicrosystemsRadio-frequency Integrated CircuitsBiomedical ApplicationsCmos Radio-frequency CircuitsFull-wave RectificationFull-wave RectifiersImplantable Bioelectronic DevicesRadio Frequency TelemetryRectifier CircuitsWireless Power Transfer For Implants
We have incorporated back telemetry mechanism within the circuitry of an integrated full-wave CMOS rectifier for RFID and implantable biomedical applications. The new rectifier saves silicon area by eliminating the need for additional large switches and provides more flexibility in choosing the most appropriate load shift keying mechanism by having the capability of both shorting and opening the transponder coil. This would result in a more robust back telemetry link with improved reading range and higher data rate. A prototype version of the rectifier was implemented in AMI 0.5-μm standard CMOS process, occupying ~0.25 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . The back telemetry rectifier was powered through an inductive link and proved to be fully functional in its three modes of operation: rectification, open coil (OC), and short coil (SC).
7
Design and analysis of an adaptive transcutaneous power telemetry for biomedical implants
Guoxing Wang, Wentai Liu, Mohanasankar Sivaprakasam et al. · IEEE Transactions on Circuits and Systems I Fundamental Theory and Applications · 2005 · 433 citations
Engineering, Biomedical Engineering, Electrical Engineering +16