Journal of Research of the National Institute of Standards and Technology · 1995 · 49 citations · 6 references
EngineeringElectrical EngineeringElectromagnetismInstrumentationMetrologyCalibrationEquipment CalibrationMagnetic SusceptibilityMass MeasurementMass MetrologyMeasurement StandardsCalibration Of Mass StandardsMagnetic PropertiesMagnetic Property MeasurementMagnetometer CalibrationMass PrecisionMass StandardsPrecision MetrologyVolume Magnetic Susceptibility
Magnetic interactions may lead to errors in precision mass metrology. An analytical description of such magnetic errors is presented in which the roles of both the volume magnetic susceptibility and permanent magnetization are discussed. The same formalism is then used to describe in detail the calibration and operation of a susceptometer developed at the Bureau International des Poids et Mesures (BIPM). The device has been optimized for the determination of the magnetic properties of 1 kg mass standards.
6
Demagnetizing factors for cylinders
D.-X. Chen, Johannes Brug, Ron Goldfarb · IEEE Transactions on Magnetics · 1991 · 516 citations · Full text
Permanent Magnets in Theory and Practice
R Carey · Physics Bulletin · 1978 · 124 citations
New method to measure magnetic susceptibility
Richard Davis · Measurement Science and Technology · 1993 · 28 citations