Journal of Separation Science · 2008 · 31 citations · 12 references
PharmacyBiochemistryAnalytical ChemistryChromatographic AnalysisSteroid BiochemistryChromatographic PrinciplesChromatographyLiquid ChromatographyMobile Phase OptimizationMobile Phase SelectionStationary Phase ChemistryAnalytical Liquid ChromatographyHigh-performance Liquid ChromatographyHigh-performance Liquid Chromatography Mobile Phase OptimizationHigh-performance Liquid Chromatography Stationary PhasesSelectivitySteroid DetectionTheoretical Plate CountTheoretical Plate Number
A new method using phase optimized LC (POPLC) for the analysis of steroids is described. The retention factors and the theoretical plate numbers of different steroids were determined for four different stationary phases. Based on these values, an optimal stationary phase composition and the resultant chromatogram have been calculated by an optimization software and compared to the measured data. It is demonstrated that these predicted data show only little differences to the experimental results. Furthermore, it is shown that the overall selectivity of the optimal stationary phase composition is much better than the selectivity of any individual stationary phase.
12