Secondary neutron dose measurement for proton eye treatment using an eye snout with a borated neutron absorber

Dong Wook Kim, Weon Kuu Chung, Jungwook Shin, Young Kyung Lim, Dongho Shin, Dong Oh Shin, Jung Keun Cho

Radiation Oncology · 2013 · 15 citations · 19 references

DOIFull text

Open access

Concepts

Abstract

BACKGROUND: We measured and assessed ways to reduce the secondary neutron dose from a system for proton eye treatment. METHODS: Proton beams of 60.30 MeV were delivered through an eye-treatment snout in passive scattering mode. Allyl diglycol carbonate (CR-39) etch detectors were used to measure the neutron dose in the external field at 0.00, 1.64, and 6.00 cm depths in a water phantom. Secondary neutron doses were measured and compared between those with and without a high-hydrogen-boron-containing block. In addition, the neutron energy and vertices distribution were obtained by using a Geant4 Monte Carlo simulation. RESULTS: The ratio of the maximum neutron dose equivalent to the proton absorbed dose (H(10)/D) at 2.00 cm from the beam field edge was 8.79 ± 1.28 mSv/Gy. The ratio of the neutron dose equivalent to the proton absorbed dose with and without a high hydrogen-boron containing block was 0.63 ± 0.06 to 1.15 ± 0.13 mSv/Gy at 2.00 cm from the edge of the field at depths of 0.00, 1.64, and 6.00 cm. CONCLUSIONS: We found that the out-of-field secondary neutron dose in proton eye treatment with an eye snout is relatively small, and it can be further reduced by installing a borated neutron absorbing material.

References

19