Improvement in clinical outcomes following optimal targeting of brain ventricular catheters with intraoperative imaging

Christopher G. Janson, Liudmila Romanova, Kyle Rudser, Stephen J. Haines

Journal of neurosurgery · 2013 · 57 citations · 16 references

Abstract

The historical target for catheter tip placement within 1 cm of the foramen of Monro in the ipsilateral lateral ventricle was associated with significantly longer revision-free survival compared with other locations. This effect remained significant after adjusting for age and whether there was a prior history of shunting. The accuracy of catheter placement in both pediatric and adult patients was strongly associated with use of intraoperative fluoroscopic confirmation. In analyses comparing intraoperative fluoroscopy and no imaging, there was a non-statistically significant difference in the 3-year time to failure, but the worst-case scenario of catastrophic short-term failure was almost completely avoided with fluoroscopy. The authors conclude that accuracy of placement is critical for shunt survival, and that use of intraoperative imaging confirmation may optimize outcomes by avoiding the majority of unacceptable placements.

References

16