British Journal of Radiology · 1996 · 77 citations · 5 references
Veterinary MedicineOral RadiologyRadiologyDental RadiologyDiagnostic ImagingDiagnostic RadiologyDental ImagingDental RadiographyDetector TechnologyImage QualityRadiation Imaging DetectorsRadiographic ImagingDetective Quantum EfficiencyDetector NoiseModulation Transfer FunctionModulation Transfer Function AnalysisPhotoinduced LuminescenceQuantum EfficiencyQuantum Efficiency MeasurementX-ray Imaging Performance
Direct digital imaging systems are becoming increasingly common in dental radiography. The majority of these systems are based on charge coupled device (CCD) technology. A relatively new system, based on photo-stimulable phosphor luminescence (PSPL), is now commercially available. This system appears to overcome some of the restrictions of CCD systems including those associated with the bulky detector, connecting wire, limited image size and limited exposure latitude. A physical evaluation of the PSPL system was conducted including measurements of sensitometric response, modulation transfer function (MTF), noise power spectrum (NPS), noise equivalent quanta (NEQ) and detective quantum efficiency (DQE). These measurements were compared with results obtained previously for Kodak Ektaspeed (E-speed) film. The results of the evaluation indicate that the PSPL system may be capable of operating at exposures up to 80% lower than for film with comparable image quality.
5
Quantitative assessment of a new dental imaging system
K. A. Homer, J Czajka, A.C. Shearer et al. · British Journal of Radiology · 1991 · 30 citations