Journal of Analytical Atomic Spectrometry · 2002 · 13 citations · 21 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisMass Spectrometric AnalysisIsotope ShiftsMass Spectrometric Elemental AnalysisMass SpectrometryRadiochemical AnalysisTrace Element AnalysisIsotope Ratio AnalysisIsotope Ratio Mass SpectrometryIsotopic Ratio DeterminationMass Bias CorrectionReference MaterialsUranium AnalysisUranium Isotopic CompositionUranium Measurement
The effect of alkali metals on the accuracy of uranium isotope ratio measurements by inductively-coupled plasma sector field mass spectrometry (ICP-SFMS) was investigated. The isotope ratio measurements were performed on the isotopic reference materials, NBS SRM U-015 and U-350, in solutions containing various amounts of any one of Na, K and Rb. The mass bias of the instrument was corrected using an isotopic standard solution without alkali metals. The observed 235U/238U isotope ratios for SRM U-015 began to increase at around 1 μg g−1 for K and Rb, and at around 10 μg g−1 for Na. The deviations from the certified value at the concentration of 300 μg g−1 were evaluated to be 0.8%, 1.9% and 2.5% for Na, K and Rb, respectively. For SRM U-350, the 234U/238U, 235U/238U and 236U/238U isotope ratios were measured in various K concentrations. The deviation per unit mass at 300 μg g−1 K was evaluated as 0.56%.
21
Yasuyuki Muramatsu, W. Rühm, Satoshi Yoshida et al. · Environmental Science & Technology · 2000 · 172 citations