IEEE Transactions on Microwave Theory and Techniques · 2008 · 619 citations · 28 references
EngineeringBiomedical EngineeringElectrical EngineeringAntennasBiosensorsMicrowave EngineeringAntenna DesignAntenna MeasurementImplantable Biomedical DevicesImplantable BiosensorsMedical Device DesignMicrowave AntennasDual-band Antenna DesignElectrical PropertiesGlucose MonitoringGlucose SensingImplantable Antenna DesignImplantable AntennasParticle Swarm OptimizationTelemetry Antenna Design
The study focuses on continuous glucose monitoring but notes that the developed antenna and skin‑mimicking gels are applicable to other wireless telemetry uses. The authors aim to design a compact dual‑band implantable antenna operating in MICS (402‑405 MHz) and ISM (2.4‑2.48 GHz) bands and to create skin‑mimicking gels for in‑vitro testing. They optimized the antenna using a finite‑element boundary‑integral electromagnetic simulation code coupled with a particle‑swarm optimization algorithm, then fabricated and measured it in the developed gel. Simulations and measurements yielded bandwidths of 20.4 % (MICS) and 4.2 % (ISM) versus 35.3 % (MICS) and 7.1 % (ISM), demonstrating satisfactory dual‑band performance.
In this study, we present a small-size dual medical implant communications service (MICS) (402-405 MHz) and industrial, scientific, and medical (ISM) (2.4-2.48 GHz) band implantable antenna for continuous glucose-monitoring applications. The antenna is optimized for dual-band operation by combining an in-house finite-element boundary integral electromagnetic simulation code and particle swarm optimization algorithm. In order to test the designed antenna in vitro, gels mimicking the electrical properties of human skin are also developed. The optimized antenna is fabricated and measured in the gel. The simulated and measured bandwidths are found to be 20.4% MICS, 4.2% ISM, and 35.3% MICS, and 7.1% ISM, respectively. Although we have emphasized continuous glucose monitoring throughout this paper, the antenna and skin mimicking gels presented here can be used for many other wireless telemetry applications.
28