Concepedia

Publication | Closed Access

Sensitivity of low energy brachytherapy Monte Carlo dose calculations to uncertainties in human tissue composition

94

Citations

46

References

2010

Year

Abstract

Low energy brachytherapy dose distributions in tissue differ from water and are influenced by density, mean tissue composition, and patient-to-patient composition variations. The results support the use of a dose calculation algorithm accounting for heterogeneities such as MC. Since this work shows that variations in mean tissue compositions affect MC dosimetry and result in increased dose uncertainties, the authors conclude that imaging tools providing more accurate estimates of elemental compositions such as dual energy CT would be beneficial.

References

YearCitations

Page 1