The new Complior Analyse device provides equivalent results for PWV and central pressure values to the Sphygmocor and Complior SP. It reaches Association for the Advancement of Medical Instrumentation standard for central blood pressure and grades as excellent for PWV on the Artery Society criteria. It can be interchanged with existing devices.