Publication | Closed Access
Comparison of the Complior Analyse device with Sphygmocor and Complior SP for pulse wave velocity and central pressure assessment
44
Citations
46
References
2014
Year
The new Complior Analyse device provides equivalent results for PWV and central pressure values to the Sphygmocor and Complior SP. It reaches Association for the Advancement of Medical Instrumentation standard for central blood pressure and grades as excellent for PWV on the Artery Society criteria. It can be interchanged with existing devices.
| Year | Citations | |
|---|---|---|
Page 1
Page 1