Journal of Chromatographic Science · 2012 · 11 citations · 14 references
Life SciencesBiochemistryEndocrinologyChromatographic AnalysisMass Spectrometric AnalysisSteroid BiochemistryMass SpectrometrySteroid Hormone BiochemistryHormonal AlterationsLinearityMass Spectrometric QuantificationMass Spectrometry-based Hormone AnalysisPlasma Steroid ProfilesPlasma Steroid ProfilingSteroid DetectionSteroid Hormone Assays
Studying the plasma steroid profile offers information about the possible existence of endocrinological alterations. This study describes the development and validation of gas chromatographic-mass spectrometric and gas tandem mass spectrometric methods for the simultaneous identification of 17 steroid hormones in human plasma using five different solvents. The n-hexane/ethyl acetate solvent mixture, in a proportion of 70/30 (v/v) provided the best results. The extracts were derivatized with N-methyl-N-trimethylsilyl-trifluoroacetamide. The obtained limits of detection were below 1 ng/mL in the majority of the studied steroids and the limits of quantification were below 5 ng/mL; the method obtained good linearity, reproducibility, repeatability, accuracy and recoveries above 95% in most cases.
14
Uta Ceglarek, Linda Kortz, Alexander Leichtle et al. · Clinica Chimica Acta · 2008 · 122 citations
Steroid measurement with LC–MS/MS in pediatric endocrinology
Manfred Rauh · Molecular and Cellular Endocrinology · 2008 · 101 citations · Full text
Medicine, Endocrinology, Pediatrics +15