Radiology · 2004 · 56 citations · 10 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyCerebral Blood FlowCerebrovascular ImagingComputed TomographyPerfusion ImagingTissue Perfusion AssessmentVascular ImagingCerebral Blood Flow QuantificationComputed Tomography PerfusionPerfusion Computed TomographyPerfusion MapsPerfusion Parameter EstimationSignal-to-noise RatioSignal-to-noise Ratio AnalysisSignal-to-noise Ratio Enhancement
The authors reviewed 40 computed tomographic (CT) perfusion studies to determine the effect of arterial and venous input function properties on perfusion parameter values and tissue signal-to-noise ratio (SNR). A 10-subject subset was analyzed to evaluate the effect of varying venous region of interest (ROI) locations. Venous peak enhancement correlated significantly with mean tissue cerebral blood flow (CBF) and cerebral blood volume (CBV); venous and arterial peak enhancement correlated significantly with SNR of perfusion maps. Different ROI locations within the same vein resulted in significantly different CBV and CBF values. Perfusion map parameters are related to peak enhancement within user-selected ROIs. Optimal ROI selection should limit variability and increase quality of CT perfusion images.
10
Perfusion CT of the brain: diagnostic approach for early detection of ischemic stroke.
M. Glenn Koenig, E Klotz, Banja Luka et al. · Radiology · 1998 · 319 citations
Dynamic CT measurement of cerebral blood flow: a validation study.
Aleksa Cenic, Darius G. Nabavi, Rosemary A. Craen et al. · PubMed · 1999 · 306 citations
Medicine, Neuroscience, Radiology +16