Exit of catheter lock solutions from double lumen acute haemodialysis catheters an in vitro study

52

Citations

9

References

2007

Year

Abstract

All double lumen dialysis catheters we tested have a substantial amount of leak even when the catheter 'lock' volumes were used, and leak ratio increases significantly with 20% overfill. There is a leak even when using 20% less 'lock' volume. The amount of leak can be clinically important and may explain the reports of bleeding episodes after heparin lock and antibiotic toxicity after antibiotic and anticoagulant combination lock. Some devices have lower leak ratios than others, likely related to catheter design.