IEEE Transactions on Applied Superconductivity · 2010 · 15 citations · 4 references
Natural SciencesMaterials EngineeringMaterials ScienceComputational Materials ScienceElectromagneticsMagnetic MaterialsFerromagnetic MaterialsMagnetic CharacterizationMagnetic Resonance ImagingMagnetic Resonance Signal FormationMagnetization ProcessesSuperconducting Magnet SystemsHigh-temperature Superconducting MagnetsMagnet TestingMagnetic Design OptimizationMagnetic Field StrengthNuclear Magnetic Resonance StudiesSpatial HomogeneitySuperconducting Magnet DesignTemporal Stability
In design and manufacture of a compact NMR (nuclear magnetic resonance) magnet which consists of a stack of high temperature superconducting (HTS) bulk annuli, spatial homogeneity and temporal stability of trapped magnetic fields are key issues. This paper presents a study on optimized configuration of stacked HTS bulk annuli for the compact NMR application. A three-stack bulk magnet was designed and tested at two different temperatures of 20 and 77 K, and two different background fields of 3 and 5 T. The test results were compared with analytic results to verify the validity of the calculation. Then, various optimized patterns were investigated numerically to find out an optimized stack pattern of HTS bulk annuli in a given condition. The results prove that with the proposed optimized method trapped field strength as well as its homogeneity of a bulk-annulus stack can be improved.
4
Takashi Nakamura, Yoshitaka Itoh, Masaaki Yoshikawa et al. · Concepts in Magnetic Resonance Part B · 2007 · 116 citations · Full text