Journal of Environmental Monitoring · 2002 · 118 citations · 14 references
Natural SciencesChemical EngineeringChemistryAir PollutionAnalytical ChemistryEnvironmental ChemistryAir Pollution InstrumentationEnvironmental AnalysisEnvironmental Analytical ChemistryTrap AnalysisVolatile Organic CompoundsVolatile Substance AnalysisFlame DetectionFlame Ionization DetectionOpen Tubular ColumnsOxygenated Volatile Organic CompoundsVolatile Organic Compound AnalysisVolatile Organic Compound DetectionVolatile Organic Compound MonitoringVolatile Organic Compounds Analysis
A method has been developed for concurrent analysis of C2-C7 hydrocarbons and C2-C5 oxygenated volatile organic compounds (o-VOCs) including alcohols, aldehydes, ketones and ethers. A multi-bed, Peltier-cooled adsorbent trap, consisting of Carboxen 1000 and Carbopack B, was used to acquire one sample per hour. Upon injection the sample was split in an approximately 50:50 ratio between a 50 m aluminium oxide (Al2O3) porous layer open tubular (PLOT) column and a 10 m LOWOX column. Eluents from each column were then analysed using flame ionisation detection (FID). Regular calibration of the system was performed using a standard cylinder mixture at the parts per billion by volume (ppbV) level for non-methane hydrocarbons (NMHCs) and a permeation tube method for the oxygenated species. The system is fully automated with NMHC detection limits between 1 and 10 parts per trillion by volume (pptV) and o-VOC detection limits between 10 and 40 pptV.
14
Richard G. Derwent, Trevor J. Davies, Michael F. Delaney et al. · Atmospheric Environment · 2000 · 206 citations
Natural Sciences, Earth Sciences, Environmental Sciences +10