Blind Monte Carlo detection-estimation method for optical coherence tomography

Georg Kail, Clemens Novak, Bernd Höfer, Franz Hlawatsch

2009 · 10 citations · 9 references

Concepts

Abstract

We consider the parametric analysis of frequency-domain optical coherence tomography (OCT) signals. A Monte Carlo (Gibbs sampler) detection-estimation method for determining the depths and reflection coefficients of tissue interfaces (reflective sites in the tissue) is proposed. Our method is blind since it estimates the instrumentation-dependent ldquofringerdquo function along with the tissue parameters. Sparsity of the detected interfaces is enforced by an impulse detector and a modified Bernoulli-Gaussian prior with a minimum distance constraint. Numerical results using synthetic and real signals demonstrate the excellent performance and fast convergence of our method.

References