Publication | Open Access
Excessive negative venous line pressures and increased arterial air bubble counts during miniaturized cardiopulmonary bypass: an experimental study comparing miniaturized with conventional perfusion systems
16
Citations
18
References
2013
Year
Despite the benefits of MCPB regarding systemic inflammatory response and haemolysis, this technique is associated with excessive negative venous line pressures and a significant increase in the number and volume of arterial air bubbles compared with CCPB. Mini-perfusion systems and the management of MCPB require further refinements to avoid such adverse effects.