IEEE Transactions on Nuclear Science · 1999 · 78 citations · 12 references
EngineeringBiomedical EngineeringNuclear MedicineBiomedical ImagingDiagnostic ImagingOpticsFiber OpticsLight-tissue InteractionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsOptical FibersPositron Emission TomographyDetector Energy ResolutionIn Vivo ImagingIn Vivo Small Animal ImagingPhotomultiplier DetectorsPhotomultiplier TubesRadiation Energy ResolutionSmall Animal ImagingStereotactic Frames
Using a combination of Monte Carlo simulations and experimental measurements we have designed a small-animal MRI-compatible PET (McPET) scanner for simultaneous PET imaging and MRI of mice and rats in vivo. The scanner consists of one ring of 480 LSO (Lu/sub 2/SiO/sub 5/) crystals arranged in three layers, with 160 crystals per layer. The crystal dimensions are 2/spl times/3/spl times/7.5 mm/sup 3/. This was based on a target resolution of 215 mm and simulations showing that a depth of 7.5 mm avoided significant depth-of-interaction effects across the desired field of view. The system diameter of 11.2 cm is large enough to accommodate the animal positioned inside a stereotactic frame. Each crystal is coupled through 2-mm diameter optical fibers to multi-channel photomultiplier tubes which reside outside the main magnetic field. Through 50 cm of optical fiber, a photopeak is clearly seen and the measured energy resolution is 25%. Prototype optical fiber connectors have been tested to increase the flexibility of the system and result in a light loss of only 6%. The proposed system has adequate resolution and sensitivity for a number of applications with small animals and is the first practical device for simultaneous in-vivo imaging with PET and MRI.
12
MicroPET: a high resolution PET scanner for imaging small animals
Simon R. Cherry, Yiping Shao, Robert W. Silverman et al. · IEEE Transactions on Nuclear Science · 1997 · 615 citations
Simultaneous PET and MR imaging
Yiping Shao, Simon R. Cherry, Keyvan Farahani et al. · Physics in Medicine and Biology · 1997 · 373 citations
Development of a PET detector system compatible with MRI/NMR systems
Yiping Shao, Simon R. Cherry, Keyvan Farahani et al. · IEEE Transactions on Nuclear Science · 1997 · 190 citations