Development and Validation of a Time-Lag Correction for Vaisala Radiosonde Humidity Measurements
326
Citations
19
References
2004
Year
This study presents a method of improving the accuracy of relative humidity (RH) measurements from Vaisala RS80 and RS90 radiosondes by applying sensor-based corrections for well-understood sources of measurement error. Laboratory measurements of the sensor time constant as a function of temperature are used to develop a correction for a time-lag error that results from slow sensor response at low temperatures. The time-lag correction is a numerical inversion algorithm that calculates the ambient (''true'') humidity profile from the measured humidity and temperature profiles, based on the sensor time constant. Existing corrections for two sources of dry bias error in RS80 humidity measurements are also included in the correction procedure: inaccuracy in the sensor calibration at low temperatures, and chemical contamination of sensors manufactured before June 2000 by nonwater molecules from the radiosonde packaging material.