Journal of Separation Science · 2012 · 13 citations · 26 references
Natural SciencesChemical EngineeringPolymer ChemistrySeparation ProcessesSeparation ScienceColumn TechnologyPolymer AnalysisPolymer CharacterizationPorous MaterialsAnalytical SeparationsCapillary ColumnsColumn Performance EvaluationColumnsMolar MassPolystyrenePore StructureSynthetic Polymers
Monolithic capillary columns based on divinylbenzene were synthesized using different alcanols as porogens. Prepared columns were tested in separation of polystyrene standards according to their molar mass (MM) and were characterized by corresponding calibration graphs. It was demonstrated that a decrease of alcanol chain length from dodecanol to octanol resulted in a decrease of column permeability and in an improved column ability to separate polystyrene standards. In contrast, removing a good solvent from porogen mixture results in an increase of column permeability and in a decrease of column performance toward polystyrene standards. Optimized synthetic conditions included porogen composed of nonanol and toluene or mesytilene, and the column prepared with this porogen was capable of separating a mixture of 14 polystyrene standards with MM ranged from several millions to oligomers.
26
Monolithic silica columns for high-efficiency chromatographic separations
Nobuo Tanaka, Hiroshi Kobayashi, N. Ishizuka et al. · Journal of Chromatography A · 2002 · 489 citations