Singly, all three neonatal hearing screening tests resulted in low refer rates, especially if referrals for follow-up were made only for the cases in which stopping criteria were not met in both ears. Following a protocol similar to that recommended in the National Institutes of Health (1993) Consensus Conference report resulted in refer rates that were less than 4%. TEOAEs at 80 dB pSPL, DPOAE at L1 = 65, L2 = 50 dB SPL and ABR at 30 dB nHL measured during the neonatal period, and as implemented in the current study, performed similarly at predicting behavioral hearing status at 8 to 12