Publication | Closed Access
Mitral valve hemodynamic effects of percutaneous edge‐to‐edge repair with the MitraClip™ device for mitral regurgitation
86
Citations
12
References
2006
Year
Echocardiographic and hemodynamic measurements after percutaneous mitral valve repair with the MitraClip show an expected decrease in mitral valve area with no evidence of clinically significant mitral stenosis either immediately after clip deployment or after 12 months of follow-up.