Clinical Chemistry · 1977 · 51 citations · 0 references
MedicineHematologyPhysiologyElectroanalytical MethodsHematology PhysiologyMineral MetabolismElectrolyte AnalysisElectrolyte BalanceElectrolyte DisordersElectrolyte HomeostasisPotassium HomeostasisRed Blood Cell PhysiologyFlame PhotometryPhotometryPotassium Concentration AnalysisPotassium Concentration MeasurementReference Range DeterminationReference Range EstablishmentReference Range StandardsSerum Electrolyte Measurement
Separated erythrocytes are washed repeatedly with iso-osmolar magnesium chloride solution, lysed by adding saponin, and sodium potassium measured in the diluted hemolysate by flame photometry. The coefficient of variation for the method was less than 4%. Reference intervals determined for a healthy population and hospitalized (elective surgery) patients without electrolyte disorders were 4.6-7.8 mmol/liter for erythrocyte sodium concentration and 94-110 mmol/liter for erythrocyte potassium concentration (2.5-97.5 percentiles).