Perioperative outcomes using LigaSure compared with conventional bipolar instruments in laparoscopic hysterectomy: a randomised controlled trial
49
Citations
28
References
2011
Year
No differences in operating time (from initial skin incision till uterine detachment and initial skin incision till final skin closure) using LigaSure versus conventional bipolar instruments: 97.6 versus 91.8 minutes (P = 0.39, 95% CI - 7.6 to 19.2), and 148.1 versus 142.1 minutes (P = 0.46, 95% CI - 10.1 to 22.3), respectively. The mean blood loss using LigaSure versus conventional bipolar was 234.1 versus 273.1 ml (P = 0.46, 95% CI -39.1 to 52.7). Various subjective efficacy and instrument handling parameters were significantly different between the two instruments and between the different participating centres. CONCLUSIONS There were no significant differences in operating time and blood loss between the use of LigaSure and the use of conventional bipolar instruments during laparoscopic hysterectomy, even after correction for potential confounders. User satisfaction parameters were assessed as significantly different by surgeons of the participating centres.