Publication | Closed Access
High-speed large-angle mammography tomosynthesis system
24
Citations
0
References
2006
Year
Computed TomographyEngineeringBiomedical EngineeringCalibrationCt ScanBreast ImagingPhoton-counting Computed TomographyInstrumentationRadiation OncologyNuclear MedicineRadiologyHealth SciencesMedical ImagingCompression PaddleUltrasoundRadiographic ImagingProjection ImagesStandard MammographyBiomedical Imaging
A new mammography tomosynthesis prototype system that acquires 21 projection images over a 60 degree angular range in approximately 8 seconds has been developed and characterized. Fast imaging sequences are facilitated by a high power tube and generator for faster delivery of the x-ray exposure and a high speed detector read-out. An enhanced a-Si/CsI flat panel digital detector provides greater DQE at low exposure, enabling tomo image sequence acquisitions at total patient dose levels between 150% and 200% of the dose of a standard mammographic view. For clinical scenarios where a single MLO tomographic acquisition per breast may replace the standard CC and MLO views, total tomosynthesis breast dose is comparable to or below the dose in standard mammography. The system supports co-registered acquisition of x-ray tomosynthesis and 3-D ultrasound data sets by incorporating an ultrasound transducer scanning system that flips into position above the compression paddle for the ultrasound exam. Initial images acquired with the system are presented.