MR imaging-related heating of deep brain stimulation electrodes: in vitro study.
127
Citations
27
References
2003
Year
A potential exists for excessive MR imaging-related heating in patients with deep brain stimulation electrodes; however, the temperature increases are linearly related to SAR values. Clinical imaging sequences that are associated with tolerable temperature elevations in the <or=2.0 degrees C range at the electrode tips can be performed safely within an SAR range <2.4 W/kg local (0.9 W/kg whole body averaged).