A non-invasive CSF flowmeter

Mitsuo Numoto, Mitsuhiro Hara, Tatsuo Sakai, Chikafusa Kadowaki, Kazuo Takeuchi

Journal of Medical Engineering & Technology · 1984 · 18 citations · 2 references

Abstract

A new non-invasive method for quantitative measurement of cerebrospinal fluid (CSF) flow in the ventriculo-peritoneal shunt tubing, used in hydrocephalus, has been developed. It is an implantable device which produces a bubble in the shunt tubing by electrolysis. This bubble is then detected in the tubing by an electrode arrangement using electric impedance or ultrasonically using a Doppler probe. The energy for electrolysis is supplied by extracorporeal high-frequency transmission. The CSF flow rate is calculated by the velocity of bubble flow in the tubing. CSF flow rates, ranging from 0.01 to 1.00 ml/min, have been measured in animal experiments with statistically good accuracy. In 11 clinical cases a flow range of between 0.01 and 1.93 ml/min have been observed.