I-GEL(TM) VS. AURAONCE(TM) LARYNGEAL MASK FOR GENERAL ANAESTHESIA WITH CONTROLLED VENTILATION IN PARALYZED PATIENTS

24

Citations

26

References

2011

Year

Abstract

First time insertions were 85.6% (i-gel) and 82% (AuraOnce) with overall success rates 96.3% (i-gel) and 94.2% (AuraOnce) (p=0.54). Average insertion times were 11.0 s (i-gel) and 11.6 s (AuraOnce) (p=0.19). Seal pressures were 30.4 cmH(2)O (i-gel) and 27.8 cmH(2)O (AuraOnce) (p=0.007). Peak pressures were 15.3 cmH(2)O (i-gel) and 15.6 cmH(2)O (AuraOnce) (p=0.57). Traumatic insertion occurred in 5.8% of i-gel™ and 2% of AuraOnce™ insertions. The overall incidence of postoperative complications was low, with the i-gel™ causing less sore throat and difficulty swallowing at 24h. Conclusion. Both devices provided effective seals for ventilation under positive pressure. I-gel™ may be a better alternative for the procedures with controlled ventilation because of higher seal pressures and lower incidence of sore throat postoperatively.