Lab on a Chip · 2007 · 76 citations · 14 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsMicroelectromechanical SystemsSensorsFlow CytometryHiv DiagnosisHiv TestingMicrofluidic DevicesMicrofluidic SystemsMicrofluidicsAccuracyCd4 Positive T CellsMicrofluidic Diagnostic DevicesMicrofluidic Diagnostic PlatformsMicrofluidic DiagnosticsMicrofluidic Flow CytometryMicrofluidic SensingOptical Detection
A novel technique is reported for counting the number and the percentage of CD4+ T lymphocytes in a polydimethylsiloxane (PDMS) microchannel. This system integrates optical fluorescence detection with resistive pulse sensing enhanced by a metal oxide semiconductor field effect transistor (MOSFET). The MOSFET signal indicates the total number of the cells passing through the detection channel, while the concurrent fluorescence signal records only the number of cells tagged with a specific fluorescent dye. The absolute count of the CD4+ T cells and its percentage to the total lymphocytes can be analyzed by combining the two counting results, which shows comparable accuracy to those from the commercial flow cytometer. The fastest observed counting rate for a single-channel microchip is 8.5 cells per second. This technique is highly promising as it could greatly reduce the cost for HIV diagnosis and treatment and make it accessible to resource-poor developing countries.
14
Current-induced magnetization reversal in nanopillars with perpendicular anisotropy
S. Mangin, D. Ravelosona, J. A. Katine et al. · Nature Materials · 2006 · 1.3K citations · Full text
DNA-Mediated Fluctuations in Ionic Current through Silicon Oxide Nanopore Channels
H. Chang, Farhad Kosari, G. Andreadakis et al. · Nano Letters · 2004 · 362 citations