Neuroradiologic Applications with Routine C-arm Flat Panel Detector CT: Evaluation of Patient Dose Measurements

Y. Kyriakou, G. Richter, Arnd Dörfler, Willi A. Kalender

American Journal of Neuroradiology · 2008 · 48 citations · 19 references

DOIFull text

Open access

Abstract

The high-speed scans used for 3D-DSA revealed a significantly lower CTDI(w) and DLP compared with clinical CT. The high-quality FPCT protocol resulted in a higher dose and should therefore be limited to acute cases, when patient transfer to a CT scanner is considered to be a disadvantage for patient management.