IEEE Transactions on Medical Imaging · 2007 · 90 citations · 21 references
MedicineRadiation OncologyRadiologyDiagnostic ImagingDiagnostic RadiologyTreatment PlanningComputed TomographyComputed Tomography PhysicsImage RegistrationMedical ImagingRadiographic ImagingThree-dimensional ReconstructionCompton Scatter EstimationComputer TomographyCt Image ReconstructionDeformation ModelsPhantom ValidationRespiratory Motion CompensationRespiratory Motion CorrectionTomographic Imaging Systems
Respiratory motion remains a significant source of errors in treatment planning for the thorax and upper abdomen. Recently, we proposed a method to estimate two-dimensional (2-D) object motion from a sequence of slowly rotating X-ray projection views, which we called deformation from orbiting views (DOVs). In this method, we model the motion as a time varying deformation of a static prior of the anatomy. We then optimize the parameters of the motion model by maximizing the similarity between the modeled and actual projection views. This paper extends the method to full three-dimensional (3-D) motion and cone-beam projection views. We address several practical issues for using a cone-beam computed tomography (CBCT) scanner that is integrated in a radiotherapy system, such as the effects of Compton scatter and the limited gantry rotation for one breathing cycle. We also present simulation and phantom results to illustrate the performance of this method.
21
William H. Press, Saul A. Teukolsky, William T. Vetterling et al. · 1994 · 15.4K citations
Respiratory correlated cone beam CT