Journal of Veterinary Emergency and Critical Care · 1996 · 14 citations · 14 references
EngineeringPhysiologyVeterinary PhysiologyAnesthesia For Small AnimalsBioinstrumentationHemodynamic ManagementCardiovascular MonitoringHemodynamicsRespiratory Monitoring InstrumentsArterial Blood Gas MonitoringArterial Oxygen SaturationAssessment Of Oxygen SaturationOxygen Desaturation TestingOxygen SaturationOxygen Saturation MeasurementPulse OximetersPulse Oximetry
Summary Ten, anesthetized dogs were instrumented with three pulse oximeter probes; two lingual transmittance probes and one rectal reflective probe. Arterial oxygen desaturation was produced by decreasing the inspired oxygen concentration. Hypotension was produced with an infusion of nitroprusside. Simultaneous pulse oximeter readings (SpO 2 ) were compared to co‐oximeter measured arterial saturation (SaO 2 ) collected over a range of SaO 2 (50–100%) and mean arterial pressures (40–100mmHg). Each of the monitors and means of evaluating SpO 2 studied provided accurate SpO 2 measurements over a range of mean arterial pressure from 40–100mmHg. All of the monitors tested tended to overestimate the SaO 2 when the arterial saturation was less than 70%.
14
A H BECKETT, M ROWLAND, P TURNER · The Lancet · 1965 · 16.9K citations
Blood Flow Limits and Pulse Oximeter Signal Detection
Dan Lawson, I. Norley, Greg Korbon et al. · Anesthesiology · 1987 · 142 citations · Full text