Validation and clinical feasibility of nitrogen washin/washout functional residual capacity measurements in children

Cecilia Olegård, Jan Pålsson, S. Lundin, O. Stenqvist

Acta Anaesthesiologica Scandinavica · 2009 · 11 citations · 22 references

Concepts

Abstract

BACKGROUND: The functional residual capacity (FRC) is an important parameter in pediatric respiratory monitoring but it is difficult to assess in the clinical setting. We have introduced a modified N(2) washout method utilizing a change of F(I)O(2) of 0.1 for FRC measurement in adult respiratory monitoring. This study validated the algorithm in a pediatric lung model and investigated the stability and feasibility in a pediatric peri-operative and intensive care setting. METHODS: The lung model was ventilated in combinations of ventilatory modes, CO(2) production, model FRC and respiratory rates. Sixteen children from 10 days to 5 years were studied peri-operatively with controlled ventilation using a Mapleson D system and in the intensive care unit using a Servo-i ventilator in a supported spontaneous mode. FRC was measured during stable metabolic, respiratory and circulatory periods at positive end expiratory pressure of 3-4 and 7-8 cmH(2)O. RESULTS: In the model and in the clinical setting, we found an excellent agreement between washout and washin measurements of FRC as well as acceptable coefficients of repeatability. CONCLUSION: FRC was satisfactorily measured by a modified N(2) algorithm and may be included as a monitoring variable in pediatric respiratory care. Pediatric FRC monitoring demands strictly stable conditions as measurements are performed close to the limits of the monitor's specifications.

References

22