American Journal of Tropical Medicine and Hygiene · 2011 · 71 citations · 8 references
MedicineHematologyGood Laboratory PracticeHematology PhysiologyTransfusion MedicineAnemiaClinical HematologyHematologic DisordersAnemia DiagnosisAnemia DiagnosticsCell Volume SensingHematological AnalysisImplementation In Low-resource SettingsScreening In Low-resource Settings
This report describes the development of a hand-powered centrifuge to determine hematocrit values in low-resource settings. A hand-powered centrifuge was constructed by using a salad spinner. Hematocrit values were measured by using the hand-powered device, and results were compared with those of a benchtop centrifuge. The packed cell volume (PCV) measured with the hand-powered device correlated linearly with results obtained with a benchtop centrifuge (r = 0.986, P < 0.001). The PCVs measured with the hand-powered centrifuge were consistently 1.14 times higher than those measured with the benchtop system. The 14% increase in PCV measured with the hand-powered centrifuge is caused by increased plasma trapped in the cell column. The reader card was adjusted to compensate for trapped plasma. A hand-powered centrifuge and calibrated reader card can be constructed for U.S. $35 and can accurately determine hematocrit values. It is suitable for use in low-resource settings because it is mechanically-powered, inexpensive, and accurate.