Sentinel lymph node detection by an optical method using scattered photons

Franklin Tellier, R. Ravelo, Hervé Simon, R. Chabrier, J. Steibel, P. Poulet

Biomedical Optics Express · 2010 · 11 citations · 19 references

DOIFull text

Open access

Concepts

Abstract

We present a new near infrared optical probe for the sentinel lymph node detection, based on the recording of scattered photons. A two wavelengths setup was developed to improve the detection threshold of an injected dye: the Patent Blue V dye. The method used consists in modulating each laser diode at a given frequency. A Fast Fourier Transform of the recorded signal separates both components. The signal amplitudes are used to compute relative Patent Blue V concentration. Results on the probe using phantoms model and small animal experimentation exhibit a sensitivity threshold of 3.2 µmol/L, which is thirty fold better than the eye visible threshold.