Concepedia

Publication | Open Access

Rapid diagnosis of enterovirus infection by a new one-step reverse transcription-PCR assay

40

Citations

15

References

1997

Year

Abstract

The AMPLICOR Enterovirus Test was evaluated with 103 cerebrospinal fluid (CSF) specimens. Twenty-seven CSF specimens were culture positive. With the AMPLICOR test, enterovirus RNA was detected in 34 specimens. Compared with culture, the AMPLICOR test gave a sensitivity of 96.3% and a specificity of 100%. The sensitivity of culture was 79.4% in comparison with the AMPLICOR test.