Simultaneous and sequential same-arm measurements in the validation studies of automated blood pressure measuring devices
11
Citations
4
References
1996
Year
HypertensionHeart FailureValidation StudiesBlood Pressure ReadingsPressure MeasurementMeasurementBlood Pressure VariabilityEngineeringWearable TechnologyDiastolic FunctionBlood PressureKinesiologyCalibrationPatient MonitoringBiostatisticsInstrumentationCardiologyBlood Flow MeasurementBlood Pressure MonitoringHeart RateSequential Same-arm MeasurementsCardiovascular DiseaseSequential MeasurementsAutomated Blood PressureMedicineMeasurement SystemEmergency MedicineAnesthesiology
The oscillometric ambulatory blood pressure recorder Daypress 500 was validated according to the British Hypertension Society protocol. Both sequential and simultaneous measurements were used. Multiple regression analysis demonstrated a significant influence of subject pulse pressure and arm circumference on device-observer systolic pressure differences. Differences between observer consecutive readings were inversely related to heart rate. Device and observer blood pressure readings were closer at simultaneous than at sequential measurements. However, both kinds of measurement led to the same final evaluation (A for diastolic and B for systolic blood pressure), provided that the appropriate grading criteria were applied for each method.