Publication | Closed Access
Determination of adequate coronary stent expansion using StentBoost, a novel fluoroscopic image processing technique
55
Citations
29
References
2006
Year
SB had superior correlations for stent expansion measured by IVUS when compared with QCA. A minimum stent diameter by SB measurement<2.5 mm is associated with inadequate stent expansion using IVUS criteria.